Y_AXIS_SETTINGS_FSV
 
 The Y_AXIS_SETTINGS_FSV node sets the y axis range. Note that these settings will not affect the trace returned by
EXTRACT_SWEEP_FSV. It will change the view on the instrument itself.
Requires a CONNECTION_FSV node at the start of the app to connect with
the instrument. The VISA address will then be listed under 'connection'.
This node should also work with compatible R&S network analyzers.  Params:    connection : VisaConnection  The VISA address (requires the CONNECTION_FSV node).   ref_level : float  The reference level (top of the y axis), in dBm.   spacing_type : select  The type of spacing for the y axis (bottom = ref - spacing).   spacing : float  Spacing for the y axis (bottom = ref - spacing), in dB or %.     Returns:    out : String  Y axis settings summary.    
   Python Code
from flojoy import flojoy, DataContainer, String, VisaConnection
from typing import Optional, Literal
@flojoy(inject_connection=True)
def Y_AXIS_SETTINGS_FSV(
    connection: VisaConnection,
    ref_level: float = -30,
    spacing_type: Literal["logarithmic", "linear %", "linear dB"] = "logarithmic",
    spacing: float = 40,
    default: Optional[DataContainer] = None,
) -> String:
    """The Y_AXIS_SETTINGS_FSV node sets the y axis range.
    Note that these settings will not affect the trace returned by
    EXTRACT_SWEEP_FSV. It will change the view on the instrument itself.
    Requires a CONNECTION_FSV node at the start of the app to connect with
    the instrument. The VISA address will then be listed under 'connection'.
    This node should also work with compatible R&S network analyzers.
    Parameters
    ----------
    connection: VisaConnection
        The VISA address (requires the CONNECTION_FSV node).
    ref_level: float
        The reference level (top of the y axis), in dBm.
    spacing_type: select
        The type of spacing for the y axis (bottom = ref - spacing).
    spacing: float
        Spacing for the y axis (bottom = ref - spacing), in dB or %.
    Returns
    -------
    String
        Y axis settings summary.
    """
    rohde = connection.get_handle()
    s = f"Y axis settings: Reference level: {ref_level}, "
    s += f"Spacing type: {spacing_type}"
    rohde.write(f"DISP:TRAC:Y:RLEV {ref_level}")
    match spacing_type:
        case "logarithmic":
            rohde.write("DISP:TRAC:Y:SPAC LOG")
        case "linear %":
            rohde.write("DISP:TRAC:Y:SPAC LIN")
        case "linear dB":
            rohde.write("DISP:TRAC:Y:SPAC LDB")
    rohde.write(f"DISP:TRAC:Y {spacing}")
    return String(s=s)
Videos
Read FM Radio Signals using Flojoy
Example
Having problems with this example app? Join our Discord community and we will help you out!
In this example, the Rohde & Schwarz FSV Signal Analyzer is used to view the FM radio range (86-110 MHz).
First 4 FSV nodes were added: CONNECTION_FSV, SWEEP_SETTINGS_FSV, INIT_SWEEP_FSV, and EXTRACT_SWEEP_FSV. A LINE node was also added. The VISA address was set for the 4 instrument nodes. The range of the signal analyzer sweep was set with SWEEP_SETTINGS_FSV (start = 86, stop = 110 MHz).
The nodes were then connected as shown in the image, and the app was run.